24 06-TTM-UM-00158 8/2012
Next><Back Cancel
File Open... File Save...
Calibration (Page 2 of 3) - General Setup
1) Please establish a
reference flow rate.
1FPS / 0.3MPS Minimum.
2) Enter the reference flow rate
below. (Do not enter 0)
3) Wait for flow to stablize.
4 Press the Set button.
Gal/MIN
Delta Time
Flow:
Export...
Edit
FIGURE 4.9 CALIBRATION PAGE 3 OF 3
Zero values are not valid for linearization entries. Flow meter
zero is entered on Page 1 of 3. If a zero calibration point is
attempted, the following error message will be shown:
Value can not be 0.
This value was already set in a previous screen (Page 1 of 3).
UltraLINK
OK
!
Press the Finish button when all points have been entered.
TARGET DBG DATA SCREEN DEFINITIONS
1) Calc Count - The number of ow calculations performed
by the meter beginning at the time the power to the meter
was last turned o and then on again.
2) Sample Count - The number of samples currently being
taken in one second.
3) Raw Delta T (ηs) - The actual amount of time it takes for
an ultrasonic pulse to cross the pipe.
4) Course Delta T - The TFX series uses two wave forms. The
coarse to nd the best delay and other timing measurements
and a ne to do the ow measurement.
5) Gain - The amount of signal ampli cation applied to the
re ected ultrasound pulse to make it readable by the digital
signal processor.
6) Gain Setting/Waveform Power - The rst number is the
gain setting on the digital pot (automatically controlled by
the AGC circuit). Valid numbers are from 1 to 100. The second
number is the power factor of the current waveform being
used. For example, “8” indicates that a ⁄ power wave form is
being used.
7) Tx Delay - The amount of time the transmitting transducer
waits for the receiving transducer to recognize an ultrasound
Target D/jointfilesconvert/1417652/bg Data
Reset
Device Type:
TFX
Tx Delay:
413
7
Flow Filter:
80
8
Sound Speed:
4900
11
Serial No (TFXD):
Calc Count:
2.2 CPS
54247
1
2
Raw Delta T (ns):
0-10.73
3
4
Gain:
66/8
430
5
6
SS (Min/Max):
OK
8.0/92.4
910
Reynolds:
0.7500
20.15
12 13
signal before the transmitter initiates another measurement
cycle.
8) Flow Filter - The current value of the adaptive lter.
9) SS (Min/Max) - The minimum and maximum signal
strength levels encountered by the meter beginning at the
time the power to the meter was last turned o and then on
again.
10) Signal Strength State - Indicates if the present signal
strength minimum and maximum are within a pre-
programed signal strength window.
11) Sound Speed - The actual sound speed being
measured by the transducers at that moment.
12) Reynolds - A number indicating how turbulent a uid
is. Reynolds numbers between 0 and 2000 are considered
laminar ow. Numbers between 2000 and 4000 are in transi-
tion between laminar and turbulent ows and numbers
greater than 4000 indicate turbulent ow.
13) Reynolds Factor - The value applied to the ow calcula-
tion to correct for variations in Reynolds numbers.
SAVING METER CONFIGURATION ON A PC
The complete con guration of the ow meter can be saved
from the Con guration screen. Select File Save button
located in the lower left-hand corner of the screen and
name the le. Files are saved as a *.dcf extension. This le
may be transferred to other ow meters or may be recalled
should the same pipe be surveyed again or multiple meters
programmed with the same information.
PRINTING A FLOW METER CONFIGURATION REPORT
Select File from the upper task bar and Print to print a cali-
bration/con guration information sheet for the installation.